Serial number
|
Fault phenomenon
|
Cause analysis
|
Treatment method
|
1
|
The buzzer sounds occasionally during the measurement
|
The workpiece is measured and judged as yesNG
|
This is a normal phenomenon and no action is required
|
2
|
Measurement interface display"Gauge requirement correction"
|
Standard partNeed calibration
|
Please calibrate in manual modeStandard part
|
3
|
There is no voltage in the separator
|
The short circuit protector works
|
Check whether the circuit is overloaded or short-circuited, and reconnect the short-circuit protector after no fault
|
4
|
The measured value of a channel of height or thickness is abnormal
|
1,AEConverter adjustment error
2,AEConverter failure
|
1readjustAEconvertor
2Replace with newAEconvertor
|
5
|
The measurement interface data does not change when the workpiece is measured
|
1,PLCwithAEConverter communication exception
2, The switching power supply has no output
|
1, inspectAEIf the cable is loose, connect the plug reliably
2Replace the switching power supply
|
6
|
The measuring fixture cannot enter the automatic measurement
|
1, did not switch to automatic mode
2, the storage Settings parameter is incorrect
|
1, switch to automatic mode
2, modify the upper and lower limits of the storage range
|
Yileride focuses on precision parts size pneumatic measurement, as a national high-tech enterprise and private science and technology enterprises in Jiangsu Province, has12Human technology research and development team, research and development personnel have participated in the United States Group, Gree, Mitsubishi, Panasonic and other companies of automatic measuring machine research and development, the company's annual research and development investment200Million, research and development of new products is not less than15PC.
If you are Billion lerideSlip sheetSPCWork station Interested or want to know more details, please click on our online customer service for consultation, or directly call the national unified service hotline 400-682-7787 . We look forward to your call!What should we pay attention to before measuring the gas momentum meter?