──Jun 02 , 2022Gas meter alarm and common fault treatment methods

Serial number

Fault phenomenon

Cause analysis

Treatment method 

1

The buzzer kept ringing during the measurement 

Workpiece measurement completed and judged asNG 

Normal phenomenon, after taking out the workpiece, the buzzer is removed  

2

The software prompts you to calibrate first

MASTER Need calibration 

Please calibrate after use 

3

The measured value of a channel is abnormal

1,AEConverter adjustment error 
2,AEConverter failure

1readjustAEconvertor 
2Replace with newAEconvertor

4

Job measuring timePCThe measurement interface data is unchanged

1,PC with AE Converter communication exception 
2, The switching power supply has no output 
3Initialization of the acquisition card failed 
4If the value is very small, the power supply of the gas-electric converter box is not connected  

1, inspectAEIf the cable is loose, connect the plug reliably 
2Replace the switching power supply 
3, OKUSBOpen the software again after the cable connection is normal 
4Verify that the power box is properly connected  

    Yileride focus on micro precision, research and development and production of digital display gas momentum meter, a variety of models, quality assurance, factory direct sales, price concessions.
    If you areDigital display gas momentum meter Interested or want to know more details, please click on our online customer service for consultation, or directly call the national unified service hotline400-682-7787. We look forward to hearing from you!


BACK
PREV

Why is it more suitable for beginners to measure workpieces with a pneumatic meter?

NEXT

Pay attention to these when measuring the digital display gas momentum meter!