──Dec 11 , 2023How to adjust the measuring magnification change and indication value of the gas momentum meter when it is unstable?

    When the measuring power of the air momentum meter changes and the indicating value is unstable, it should be adjusted as follows:
    1, magnification nailOThe ring is worn and needs to be replacedOType rubber ring;
    2The end face of the thimble is worn and dented, and the chromium oxide is ground on the grinding plate, and the flatness of the thimble is checked with a knife edge flat rule0.002μm;
    3The measuring head wear causes the magnification to decrease, and the measuring head should be replaced;
    4The lock nut and the magnification nail are adjusted to the appropriate position to achieve no obvious movement and swing.